Features
- Drive Med Aire Bariatric Alternating Pressure and Low Air Loss Mattress Replacement System comes with EZ Lock/Quick Release Tubing Connector
- Cell-on-cell design of 10" deep air cells prevents bottoming out and also provides up to 24 hours of power outage protection
- Whisper quiet pump produces 12-LPM (Litres per Minute) in air flow
- Each one of the 20, 10" deep air bladders are easily removed and replaced
- Stretch cover is low shear, fluid-resistant, vapor permeable, quilted and zippered
- 3 diagnostic alarms tell you whether a problem is with the mattress or the pump for more efficient servicing
- Pillow function maintains the 3 air cells at the head of the bed in static mode for patient comfort
- CPR (Cardiopulmonary Resuscitation) valve allows for rapid deflation
More Information
- Handling and storage:
- Lay the mattress out flat and upside down
- Roll from the foot end towards the head end, the foot-end strap can then be stretched around the rolled mattress to prevent unrolling
- Do not fold, crease or stack the mattress - Maintenance:
- Check the power cord and plug to see if there are abrasions or excessive wear
- Check the mattress cover for signs of wear or damage
- Ensure the mattress cover and tubes are connected together correctly
- Plug in the pump and check the airflow from the hose connection port
- The airflow should alternate between ports every half-cycle time
- Check the air hoses to see if there are kinks or breaks
- Make sure the mattress tube is well connected
- Check the pump and make sure both power and power indicator are off when the switch is turned off - Low pressure examination:
- Examine if there is air leakage between the pump and the mattress connections or from the air mattress tubes
- Check connectors between the air mattress and pump
- If there is any disconnection, please reconnect it
- Check the CPR valves and ensure their outlets are sealed
- Check the air-connecting tubes
- Turn the pressure-adjust knob to Firm
- Keep the tubes fully inflated and inspect for air leakage
- Check if there is any air leakage from cells
- Ensure no leakage occurs
WARNING: Cancer and Reproductive Harm -
www.P65Warnings.ca.gov
Specification
Particulars | Specifications |
Control Unit Airflow | 12 LPM (Litres per Minute) |
Control Unit Cycle Time | 10, 15, 20, 25 minutes |
Control Unit Type | Compressor |
Control Unit Alarms | Visual/Audible |
Mattress Cover Material | Fluid Resistant, 4-way Stretch |
Weight Capacity | 600lb to 1000lb |