Answer: As per manufacturer, the RAM cannula has a standard 15mm critical diameter to fit all single limb circuits. They recommend up to 4 liters, anything beyond that is up to each facility. As per Vermont Oxford definition High Flow is any LPM over 1 & will vary per hospital protocol. For heated humidification you can hook the RAM-NC directly to the inspiratory limb of the circuit coming directly off the humidifier chamber to the RAM’s inner diameter of connector interface (not all neonatal circuits will do this) with the temp probe right behind it and flow rate as needed.