Answer: Cleaning instructions
It is advisable to clean the SSV daily. Do not clean the valve
while it is still attached to the tracheostomy tube. Once the
valve has been removed from the tube, it may be cleaned with
hydrogen peroxide or with soap and water for 10 minutes. The
hydrogen peroxide should be completely flushed off with tap
water when done cleaning. The valve may also be cleaned in
any proteolytic enzyme cleaner (e.g. contact lens solution). The
SSV can be cleaned with a cotton swab without fear of damage.
Dry thoroughly after cleaning is complete. It is recommended
to have a second SSV available, so that the second SSV can be
used while the first SSV is being cleaned.
WARNING: Oniy use the cleaning solutions identified above. Do
not use a sharp object to assist in cleaning. Do not leave the
valve in hydrogen peroxide for more than 10 minutes, as doing
so can decrease the life expectancy of the valve.
When to replace/change the valve
It is advisable to replace each SSV every 2 months or 60 days.
The SSV has a life expectancy of a maximum of 3 months or 90
days of use. Discard the valve after at most 90 days of use.
Replace the valve either 90 days after it was ?rst put into use or
when any of the following occur prior to the 90-day mark: 1)
there is a noticeable change in color of the valve or the inside
ball, 2) there is an increase in the amount of thick mucus that
has adhered to the ball or the inside of the valve and that
cannot be removed with cleaning, or 3) there are any visual
indentations to the valve body or any changes in its shape
(caused by accident, misuse, etc.).