DAR Hygrobac S Filter OR Heat and Moisture Exchanger
DAR Hygrobac S Filter OR Heat and Moisture Exchanger

DAR Hygrobac S Filter OR Heat and Moisture Exchanger

Brand/Manufacturer: COVIDIEN/MEDTRONIC
Save

DAR Hygrobac S Filter OR Heat and Moisture Exchanger reduces set-up time, minimizes dead space and weight on endotracheal tube and circuit while maximizing efficiency. Electrostatic hydrophobic membrane protects against cross-contamination and hygroscopic membrane traps patients moisture providing effective airway humidification. It features low resistance to airflow to minimize work of breathing.

Item #DescPkgRewardPrice 
352U5877 22M/15F-22F/15M Connection 50/Case $2.16
$299.99
$215.99

Features

  • DAR Hygrobac S Filter OR Heat and Moisture Exchanger has effective moisture control which reduces the risk of colonization in breathing tubes and minimizes sampling line blockage
  • End tidal CO2 sampling port offers convenient access to airway gases
  • ISO standard 15mm and 22mm fitting connects with breathing system
  • Treated with physiologically safe calcium chloride
  • Supplied clean and individually packaged
  • Indications for usage:
    - Adult (recommended tidal volume >150ml), single patient use on anesthetized patients and respiratory care patients who require a breathing circuit
    - Intended to be replaced at least once every 24 hours

Specification

ParticularsDimensions
Resistance to Flow (ISO 9360)

30 l/min : 1.0cm H2O

60 l/min :  2.8cm H2O

90 l/min : 4.7cm H2O

Bacterial Filtration Efficiency>99.99%
Viral Filtration Efficiency>99.99%
Dead space49ml
Weight28gm
Type of FiltrationElectrostatic
Recommended Tidal Volume>150ml
Moisture output and temperature at 2 hours (ISO 9360)

Vt 250ml : 34.4mg H2O/l @32.5oC

Vt 500ml : 33.6mg H2O/l @ 32.3oC

Vt 750ml : 30.1mg H2O/l @ 30.6oC

Vt 1000ml : 32.9mg H2O/l @ 32.2oC

Customer Reviews

Questions & Answers


 
 
question image
DOES THIS DEVICE ELIMINATE THE NEED FOR HUMIDIFICATION ON TRACH PATIENTS AND FOR HOW LONG AND IS IT COMPATABLE WITH ALL TRACHS ...INCLUDING THE FIVE JACKSON ?
LOUIS SANTONI - 2/6/2015 8:26:43 PM
facebooktwittergplus
answer image
HMEs, or Heat and Moisture Exchangers, capture heat and water vapor from a patient’s exhaled air. Then during inhalation, they add heat and moisture to the inspired air, providing the appropriate humidification.