Instructions:
- Place the non-slip Posey Grip on the mattress
- Position sensor on top of mattress, across width of bed, in one of two positions: directly under patient’s shoulders; alarm will sound when patient sits up or directly under patient’s buttocks; alarm will sound when patient moves off the sensor as in a bed exit
- Use the metal clips to secure the sensor to the mattress
- Place the linens on the bed over the sensor
- Connect the sensor to the Posey alarm unit
- Make sure the sensor cord is not stretched or stressed in any way and is clear from entanglement in the bed frame
- Cleaning: Wipe clean with liquid disinfectant. Do not submerge in liquid or sterilize with heat
Warning:
- Read sensor and alarm instructions prior to use
- Never connect Posey sensors to other manufacturers’ alarm units
- Always test system per instructions prior to use
- Be sure restraint straps do not cross over sensor pad. Restraint straps may prevent alarm activation if patient falls off mattress
- When using a restraint, always use side rails in the “up” position
- The patient’s body or limb should not fit over, under, around, through or between side rails
- The alarm may fail to sound if the pad is bent or folded
|